Measuring range |
|
TS9100M |
TS9100W |
| Range 1 |
-20 to 100°C
(calibrated at 0
to 100°C.
outside is
theoretical value)
|
-40 to 120°C
(calibrated at 0
to 120°C.
outside is
theoretical value)
|
| Range 2 |
0 to 250°C
(calibrated at 0
to 250°C)
|
0 to 500°C
(calibrated at 0
to 500°C)
|
| Optional |
200 to 1000°C or 200 to 2000°C
(without range 1&2) |
|
Resolution
(Defined as
waveform noise/10)
|
Range 1 |
0.06°C (at 30°C 60Hz)
0.03°C (at 30°C
Σ16)
|
0.08°C (at 30°C 60Hz)
0.04°C (at 30°C
Σ16)
|
| Range 2 |
0.15°C (at 30°C 60Hz)
0.08°C (at 30°C
Σ16)
|
0.30°C (at 30°C 60Hz)
0.15°C (at 30°C
Σ16)
|
|
Accuracy |
±2°C or ±2% of reading
whichever is greater
(When ambient
temp 0-40°C, standard 21.7° lens used, focusing range 50cm)
|
|
Detector |
Uncooled focal plane array (microbolometer) |
|
Spectral range |
8 to
14µm |
|
I.F.O.V |
1.2mrad (When standard 21.7° lens is used.) |
|
Focusing range |
30cm
to infinity (motor driven only, at std, wide-angle, 29° lenses) |
|
Field of view |
Σ2,
Σ8, Σ16, and spatial
filter ON/OFF |
|
Frame time |
60
frames/sec |
|
Thermal image pixels |
320
(H) x 240 (V) pixels (Three lines on all edges are out of
specification) |
|
A/D resoution |
14bits |
|
Alarm output |
Contact output (photo MOS relay), voltage: DC +10 to 27V, current:
40mA or less |
|
Measuring functions |
Run/Freeze |
|
S/N improvement |
|
|
Event output |
Screen
display |
|
Env. temp.correction |
Provided (including interval NUC) |
|
Background compensation |
Provided |
|
Auto functions |
Full
automatic (level, sense, focus), auto level/sense (manual
available), Auto focus, level trace, auto-gain control |
|
Display functions |
Display colour:
colour/monochrome, positive/negative
Graduation: 16, 32, 62, 128,
256
Colour palette: rainbow,
brightness, shine, hot-iron, medical, fine
Isothermal band display: max 4
bands
Multi-sense display
Wave form display: X-Y waveform
at freeze mode
Alarm display: alarms of temp
and TS91000 self diagnosis
Multilingual menu
|
|
Image processing functions |
Variable level/sense (at Freeze
mode)
Multi-point temperature display
(10 points)
Multi-point emissivity display
(10 points)
t display
Max/Min (peak hold) temperature
display (full screen or specified area)
Digital zoom: 2, 4 times
(Run/Freeze)
Box setting (max. 5 boxes)
Temperature alarm output
|
|
Data display |
Colour
bar (greyscale), temp range, temp level emissivity, multi-point temp
display, time, annotation, error display, multilingual menu:
English, French, German, Portuguese, Spanish, Japanese |
|
Annotation |
Text
annotation |
|
Video signal output |
NTSC/PAL composite video signal (BNC connector) |
|
Interface |
RS-232C (Standard) IEEE1394 or Ethernet (Option: select when order) |
|
Operating temp/humidity |
-15 to
50°C, 90% RH or less (not condensed) |
|
Storage temp/humidity |
-40 to
70°C, 90% RH or less (not condensed) |
|
Power supply |
DC 11
to 13V |
|
Power consumption |
Approx. 5W(typical): at environmental temp 20°C, RUN mode, RS-232C
interface) |
|
Shock and vibration |
294m/s²
(IEC60068-2-27), 29.4m/s² (IEC60068-2-6) |
|
Environmental Protection |
IP54
(IEC60529) |
|
Dimensions |
Approx. 99(W) x 112(H) x 206(D)
mm
(w/ standard, wide-angle or 29°
lens: excluding projection)
Approx. 112(W) x 112(H) x
230(D) mm
(with 2 x telephoto lens:
excluding projection)
|
|
Weight |
Approx. 2.6Kg (w/ standard, wide-angle or 29° lens) |
|
Standard accessories |
Lens