NEC TH9260 Infrared Thermal Imaging Camera

  TH9260
Measuring range
Range 1
Range 2
Range 3
Range 4
-40 to 500°C
-20 to 60°C
-40 to 120°C
0 to 500°C
200 to 2000°C (Optional)
Resolution
Range 1: 0.06°C or better at 30°C (30Hz)
Range 2: 0.08°C or better at 30°C (30Hz)
Range 3: 0.12°C or better at 30°C (30Hz)
Accuracy
±2°C or ±2% of reading whichever is greater
Detector Uncooled focal plane array (microbolometer)
Spectral range 8 to 14µm
I.F.O.V 0.6mrad
Focusing range 30cm to infinity
Field of view 21.7° (H) x 16.4° (V)
Display View finder & 5.6" movable colour LCD
Frame time 30 frames/sec
Thermal image pixels 640 (H) x 480 (V) pixels
Emissivity Correction Provided (0.10 to 1.00)
Ambient compensation Provided
Background compensation Provided
Auto functions Full automatic (level, sense, focus)
Isothermal band display Provided (up to 4 bands)
Image processing functions
Thermal / visual image fusion display
Multi-point temperature display (up to 10 points)
BOX setting, up to 5 boxes (max, min, average)
Temperature difference between 2 points (t)
Multi-focus, Sharpness filter, Median filter
Visual Camera 1.3M pixels, colour
Video signal output NTSC/PAL, composite video signal, S-video
Interfaces
USB 2.0
IEEE1394
Compact flash memory card slot
Laser pointer Class 2 (1mW or lower, red)
Operating temperature -15 to 50°C
Power supply AC adapter: 100 to 240V AC, 7.2V DC
Shock and vibration 294m/s² (30G) (IEC60068-2-27), 29.4m/s² (3G) (IEC60068-2-6)
Environmental Protection IP54 (IEC60529)
Dimensions
110 (W) x 110 (H) x 210 (D) mm
(excluding projection)
Weight 1.7Kg (inc. battery)
Standard accessories AC adapter, battery pack (2pcs), battery charger, CF memory card (1GB), CF memory card adapter, USB cable, lens cap, grip belt, neck strap, carrying case, viewer software, operation manual, target illuminator
Viewer software
Thermal image display:
  • Thermal Image thumbnail (Windows Explorer)
  • Thermal image replay
  • Image preview
Setup & functions:
  • Level, sense, span, voice replay, selection of thermal or visual image, colour bars, graduation, page
Edit:
  • Image save (BMP or JPEG)
  • Select folder
*1 Calibration range 0 to 100°C, range other than the calibration range is theoretical value

*2 Calibration range 0 to 250°C, range other than the calibration range is theoretical value

Specifications are subject to change without prior notice.